What is the loading capacity of a reversed phase preparative column?

The loading capacity of reversed phase column depends on the property of your analytes as well as the separation you can get. Normally we suggest loading 1% of the sorbent mass. Please see the below chart for more detail.

  10 mm ID 21.2 mm ID 30 mm ID
Length (mm) Column Bed Mass 1% specific load Column Bed Mass 1% specific load Column Bed Mass 1% specific load
50 2.4g 24mg 10.6g 106mg 21.2g 212mg
75 3.5g 35mg 15.9g 159mg 31.8g 318mg
100 4.7g 47mg 21.2g 212mg 42.4g 424mg
150 7.1g 71mg 31.7g 317mg 63.6g 636mg
250 11.8g 118mg 52.9g 529mg 106g 1060mg