You can extract nucleotides from aqueous-based samples using Strata™-X-AW. Nucleotides will retain in the sorbent via weak anion-exchange interactions while you are able to apply a water and methanol wash to remove unwanted matrix interferences. This method is outlined below.
Condition/Equilibrate with 300 µL methanol, and then deionized water
Load sample slowly
(If sample is viscous, please dilute 1:1 with a pH 6 buffer)
Wash with 300 µL deionized water and then 300 µL methanol
Dry cartridge on full vacuum for 10 minutes
Elute with 2x 300 µL ammonium hydroxide/methanol (5:95)