How can I dilute samples in the EZ:faast protocol?

 

Diluting samples in the EZ:faast kit protocol is useful if concentration is outside of the standard curve (i.e. 200 nmol/mL).  The optimal way is to dilute before, e.g. process 50 μL of sample instead of 100 μL.

Alternatively, the following can be performed:

  • For GC kits, change split ratio instead of 15:1,30:1,40:1
  • Dilute with dissolution solution for MS, though this will also decrease internal standard (so only if 4x)
  • For LC/MS, inject 1 μL instead of 2 μL

 